Abstract

[Image: see text] With the global escalation of concerns surrounding prostate cancer (PCa) diagnosis, reliance on the serologic prostate-specific antigen (PSA) test remains the primary approach.

However, the imperative for early PCa diagnosis necessitates more effective, accurate, and rapid diagnostic point-of-care (POC) devices to enhance the result reliability and minimize disease-related complications.

Among POC approaches, electrochemical biosensors, known for their amenability and miniaturization capabilities, have emerged as promising candidates.

In this study, we developed an impedimetric sensing platform to detect urinary zinc (UZn) in both artificial and clinical urine samples.

Our approach lies in integrating label-free impedimetric sensing and the introduction of porosity through surface modification techniques.

Leveraging a cellulose acetate/reduced graphene oxide composite, our sensor’s recognition layer is engineered to exhibit enhanced porosity, critical for improving the sensitivity, capture, and interaction with UZn.

The sensitivity is further amplified by incorporating zincon as an external dopant, establishing highly effective recognition sites.

Our sensor demonstrates a limit of detection of 7.33 ng/mL in the 0.1–1000 ng/mL dynamic range, which aligns with the reference benchmark samples from clinical biochemistry.

Our sensor results are comparable with the results of inductively coupled plasma mass spectrometry (ICP-MS) where a notable correlation of 0.991 is achieved.

To validate our sensor in a real-life scenario, tests were performed on human urine samples from patients being investigated for prostate cancer.

Testing clinical urine samples using our sensing platform and ICP-MS produced highly comparable results.

A linear correlation with R(2) = 0.964 with no significant difference between two groups (p-value = 0.936) was found, thus confirming the reliability of our sensing platform.

Dehghani, Parisa,Karthikeyan, Vaithinathan,Tajabadi, Ataollah,Assi, Dani S.,Catchpole, Anthony,Wadsworth, John,Leung, Hing Y.,Roy, Vellaisamy A. L., 2024, Rapid Near-Patient Impedimetric Sensing Platform for Prostate Cancer Diagnosis, American Chemical Society

Document

Open Open

Share

Source

Articles recommended by ES/IODE AI

A Novel MR Imaging Sequence of 3D-ZOOMit Real Inversion-Recovery Imaging Improves Endolymphatic Hydrops Detection in Patients with Ménière Disease
ménière disease p < detection imaging sequences 3d-zoomit 3d endolymphatic real tse reconstruction ir inversion-recovery hydrops ratio
Successful omental flap coverage repair of a rectovaginal fistula after low anterior resection: a case report
rectovaginal fistula rectal cancer low anterior resection omental flap muscle flap rectal cancer pod initial repair rvf flap omental lar coverage