Dokumentdetails
ID

oai:arXiv.org:2407.14243

Thema
Astrophysics - Instrumentation and... High Energy Physics - Experiment
Autor
Yi, Difan Liu, Qian Liu, Hongbang Xie, Fei Feng, Huanbo Feng, Zuke Li, Jin Liang, Enwei Zheng, Yangheng
Kategorie

Wissenschaften: Astrophysik

Jahr

2024

Auflistungsdatum

24.07.2024

Schlüsselwörter
degree systematic modulation
Metrisch

Zusammenfassung

Gaseous X-ray polarimetry refers to a class of detectors used for measuring the polarization of soft X-rays.

The systematic effects of such detectors introduce residual modulation, leading to systematic biases in the polarization detection results of the source.

This paper discusses the systematic effects and their calibration and correction using the Gas Microchannel Plate-Pixel Detector (GMPD) prototype for POLAR-2/Low-Energy X-ray Polarization Detector (LPD).

Additionally, we propose an algorithm that combines parameterization with Monte Carlo simulation and Bayesian iteration to eliminate residual modulation.

The residual modulation after data correction at different energy points has been reduced to below 1%, and a good linear relationship is observed between the polarization degree and modulation degree.

The improvement in modulation degree after correction ranges from 2% to 15%, and the results exceed those of the Imaging X-Ray Polarimetry Explorer (IXPE) above 5 keV.